2.1 OCULAR MICROMETER
INTRODUCTION
Ocular means of or connected with eyes or vision, and a lens
is a piece of glass with curved sides that concentrates light rays and
magnifies images. Thus the ocular lens quite simply is the one you look
through. It causes the contents of the slide to appear magnified. The
curvature of the lens determines the magnification level. On the
microscope the ocular lens is found at the top of the tube which
connects it to the objectives lens or lenses.Most ocular lenses have either a 10X (magnified ten times)
or 15X magnification level. When a person looks through the ocular lens,
light reflects from the objective lens, and the slide is magnified by
the multiplied magnification levels of both the ocular and the objective
lenses. So if a person is using a microscope with an ocular lens of 10X
and they set the objective lens to 4X, then the slide will appear 40X
when viewed through the ocular lens. Most microscopes have several
settings for objective lenses.
OBJECTIVE
To measure and count cells using ocular microscope.
RESULT
400x magnification |
1000x magnification
|
|||
calibrate |
Stage micrometer |
Ocular division |
Stage
micrometer
|
Ocular
divison
|
0.03
mm
|
39
division
|
0.09
mm
|
96
division
|
|
7.69
x10⁻⁴ mm
|
1
division
|
9.38x10⁻⁴ mm
|
1
division
|
|
0.769
µm
|
1
division
|
0.94
µm
|
1
division
|
a ) Yeast
400x magnification
4 division x 0.769 µm
=3.076 µm
1000x magnification
3 division x 0.94 µm
= 2.82 µm
b ) Lactobacillus
400x magnification
4 division x 0.769 µm
=3.076 µm
1000x magnification
5
division x 0.94 µm
=4.7 µm
DISCUSSION
|
CONCLUSION
Basically, calibration is the "zero" setting on a measuring object. For
example a weighing scale that is incorrectly calibrated would, for
example, read 50 pounds as 49 pounds. This means that whatever is
actually on the scale would be one pound heavier than what it actually
is. This is not a problem when comparing two things of different weight
(the difference will be the same) but if we measuring something as a
single unit, it could be troublesome.
An ocular micrometer is what's on microscopes and basically tells
how big an object is. An incorrectly calibrated micrometer would be
catastrophic since a small difference in the reading could have a
massive effect on what we're trying to do. For example, a 2mm
difference could mistakenly read a 300% increase in bacteria growth.
Bottom line is that calibration is important in measuring results and an incorrect calibration on an instrument would yeild incorrect results.
REFERENCES
Lodish, H., et al. (2000). Molecular Cell Biology (4th ed.). New York: W.H. Freeman and Co.
Wolfe, S.L. (1993). Molecular and Cellular Biology. Belmont, CA: Wadsworth Publishing Company.
aprette, D. (1995). Light Microscopy. Retrieved 8-22-2006 from http://www.ruf.rice.edu/~bioslabs/methods/microscopy/microscopy.html
http://www.mecanusa.com/microscope/micrometer/micrometer-use_en.html
2.2 NEUBAUER CHAMBER
INTRODUCTION
Counting chamber: This one is called the Neubauer improved. There are other standards with different grids available as well. |
The hemocytometer (or haemocytometer or counting chamber) is a specimen slide which is used to determine the concentration of cells in a liquid sample. It is frequently used to determine the concentration of blood cells (hence the name “hemo-”) but also the concentration of sperm cells in a sample. The cover glass, which is placed on the sample, does not simply float on the liquid, but is held in place at a specified height (usually 0.1mm). Additionally, a grid is etched into the glass of the hemocytometer. This grid, an arrangement of squares of different sizes, allows for an easy counting of cells. This way it is possible to determine the number of cells in a specified volume.
OBJECTIVE
To find the number of cells in 1ml of original solution.
RESULT
CALCULATION :
Cell in 10 boxes = 39
Average of cell: 39 ÷ 10 = 3.9
Volume
= 0.02 x 0.02 x 0.01 mm
= 4 x 10⁻3 mm
= 4 x 10⁻3 mm
= 4 x 10⁻6 cm
= 4 x 10-6 ml
= 4 x 10-6 ml
1ml
= 3.9 ÷ (4 x 10-6 )
= 9.75 x 10-5 cells/ml
DISCUSSION
Both the standard (incline) and the new “V-Load” counting hambers for different charging methods. The tolerances are ( 2% of the volume). Cell Depth: 0.100 mm ( +/-2%); Volume: 0.1 Microliter Ruling; Pattern: Improved Neubauer, 1/400 Square mm Rulings cover 9 square millimeters. Boundary lines of the Neubauer rulings are the center lines of the groups of three. (These are indicated in the illustration ). The central square millimeter is ruled into 25 groups of 16 small squares, each group separated by triple lines, the middle one of which is the boundary. The ruled surface is 0.10 mm below the cover glass, so that the volume over each of 16.
To get an accurate result ,the fluid containing the cells must be appropriately prepared before applying it to the hemocytometer.
- Proper mixing: The fluid should be a homogenous suspension. Cells that stick together in clumps are difficult to count and they are not evenly distributed.
- Appropriate concentration: The concentration of the cells should neither be too high or too low. If the concentration is too high, then the cells overlap and are difficult to count. A low concentration of only a few cells per square results in a higher statistical error and it is then necessary to count more squares (which takes time). Suspensions that have a too high concentration should be diluted 1:10, 1:100 and 1:1000. A 1:10 dilution can be made by taking 1 part of the sample and mixing it with 9 parts water (or better saline of correct concentration to prevent bursting of the cells). The dilution must later be considered when calculating the final concentration.
- Counting cells that are on a line: Cells that are on the line of a grid require special attention. Cells that touch the top and right lines of a square should not be counted, cells on the bottom and left side should be counted.
- Number of squares to count: The lower the concentration, the more squares should be counted. Otherwise one introduces statistical errors. How many squares? To find out one could calculate the cell concentration per ml based on the numbers obtained from 2 different squares. If the final result is very different, then this can be an indication of sampling error.
- Averaging: If one did not count all of the cells in a large square (1mmx1mm) then it is necessary to average the results first before proceeding.
- Computing the volume: It is necessary to determine the volume represented by the square. The width and height of the square (e.g. 0.2mm x 0.2mm) must be multiplied by the height of the sample : v = 0.2mm x 0.2mm x 0.1mm = 0.004mm³ =0.000004 mL
- Calculating the number of cells in 1 ml
- Correcting for dilution: If the sample was diluted before counting, then this must be taking into consideration as well. We assume that the sample was diluted 1:10.
CONCLUSION
There are different types of counting chambers available, with different grid sizes. One counting chamber also has grids of different sizes. Take care that the grid size and height (read the instruction manual) otherwise we'll make calculation errors.We need to use cover glasses so that the height of the fluid is standardized.They are thicker than the standard 0.15mm cover glasses. They are therefore less flexible and the surface tension of the fluid will not deform them. For moving cells (such as sperm cells) are difficult to count. These cells must first be immobilized.
REFERENCES
http://www.emsdiasum.com
http://www.microbehunter.com
http://www.brand.de/index.php?id=497&L=0
No comments:
Post a Comment